Study: Even Jacobi-Fourier IOL Achieves >1.5 D Depth of Focus Across Pupil Sizes

An intraocular lens designed with an even Jacobi-Fourier phase profile maintained acceptable image quality across a defocus range greater than 1.5 diopters for all pupil sizes tested, and at least 2.0 diopters for pupils of 4.0 mm or smaller, according to findings published in Scientific Reports.

Researchers evaluated the lens inside a Liou-Brennan model eye using polychromatic light. Entrance pupil diameters ranged from 2.5 mm to 5.0 mm. Performance metrics included Optical Transfer Function, simulated visual acuity based on area under the OTF, simulated Sloan letter appearance, and Hilbert Space Angle.

Excepting HSA, all metrics determined the range of object vergences where image quality remained acceptable, defined as logMAR ≤ 0.2. The HSA results confirmed image quality remained steady within this range.

As a proof of concept, a JFP phase plate was fabricated and used to photograph letter charts at different object vergences. The experimental findings demonstrated that the Jacobi-Fourier phase can be successfully applied in a defocus-resistant imaging system.

Preliminary dysphotopsia simulations indicated the IOL performed acceptably.

Scientific Reports, the Nature Portfolio open-access mega journal established in 2011, publishes papers judged solely on scientific validity rather than perceived impact. It became the world’s largest journal by article count in September 2016, overtaking PLOS One. In 2026 PubMed halted indexing of new Scientific Reports papers pending review by the US National Library of Medicine.

Extended depth of focus lenses aim to reduce dependence on spectacles after cataract surgery. Conventional diffractive or refractive designs often trade contrast for range. The even Jacobi-Fourier phase approach seeks to flatten the through-focus performance curve without increasing higher-order aberrations that drive dysphotopsia.

What defocus range did the lens achieve in the model eye?
All pupil sizes from 2.5 mm to 5.0 mm showed a range greater than 1.5 D. For pupils of 4.0 mm or smaller, the range reached at least 2.0 D.
How was image quality measured?
Metrics included Optical Transfer Function, simulated visual acuity (area under OTF), simulated Sloan letters, and Hilbert Space Angle. Acceptable quality was defined as logMAR ≤ 0.2.
Was the design tested physically?
Yes. A fabricated JFP phase plate photographed letter charts at multiple vergences, confirming the defocus-resistant behavior experimentally.
Did the lens show dysphotopsia risk?
Preliminary simulations indicated acceptable dysphotopsia performance.

Did you know? The Liou-Brennan model eye includes aspheric surfaces and a gradient-index lens, making it a closer match to human optical performance than simpler four-surface models.